Ni usb-6001/6002/6003 User Guide National Instruments


+5 V Source—Regulated 5 V supply. • +5 V Protection



tải về 1.5 Mb.
Chế độ xem pdf
trang18/24
Chuyển đổi dữ liệu13.04.2022
Kích1.5 Mb.
#51632
1   ...   14   15   16   17   18   19   20   21   ...   24
NI6001
code vf-s11, BÁO CÁO MÔN ĐƯỜNG LỐI, bao cao mon hoc trang bị
+5 V Source—Regulated 5 V supply.



+5 V Protection—Circuit for overvoltage, over current, and short circuit protection.

The +5 V source is limited at 200 mA typically. In case of hard short circuit to ground, this limit 

is further reduced to avoid excessive power dissipation.

+5 V


Protection

16-Position

Screw Terminal

Plug


+5 V

D GND


I/O Connector

To USB


Connector

+5 V


Source


24

|

ni.com



|

NI USB-6001/6002/6003 User Guide

Connecting the Load

The return terminal for the +5 V source is the D GND terminal. +5 V load should be connected 

between the +5 V terminal and D GND. The current delivered by the NI DAQ device at the +5 V 

terminal is sourced from the USB connector. To meet the USB specifications, a maximum of 

150 mA can be used from the +5 V terminal.

Figure 16.  

Connecting +5 V Power Source Load

Enable +5 V Power Source 

The + 5 V power is always enabled and the voltage is present at the +5 V terminal when the 

device is in active mode. The +5 V power source is not available for about one second after the 

NI DAQ device is first connected to the USB connector or when the device is in suspend mode.

Related Documentation

The most recent versions of documents are available from 

ni.com/manuals

.

The following references to documents assume you have NI-DAQmx 9.9 or later and, where 



applicable, version 8.5 or later of the NI application software.

Examples


To locate NI software examples, go to 

ni.com/info

 and enter the Info Code 

daqmxexp


.

NI USB-6001/6002/6003

The NI USB-6001/6002/6003 Quick Start packaged with your NI DAQ device describes how to 

install the device and confirm that your device is operating properly.

The NI USB-6001 Specifications, NI USB-6002 Specifications, and NI USB-6003 Specifications 

are available for download from 

ni.com/manuals

.

16-Position



Screw Terminal

Plug


Optional

Capacitor

Load

D GND


+5 V


NI USB-6001/6002/6003 User Guide

|

© National Instruments



|

25

NI-DAQmx



The NI-DAQmx Readme lists which devices, Application Development Environments (ADEs), 

and NI application software are supported by your version of NI-DAQmx. To access the 



NI-DAQmx Readme, select Start»All Programs»National Instruments»NI-DAQmx»

NI-DAQmx Readme

The NI-DAQmx Help contains API and NI-DAQmx concepts, and applications. To access the 



NI-DAQmx Help, select Start»All Programs»National Instruments»NI-DAQmx»

NI-DAQmx Help

LabVIEW


If you are a new user, go to 

ni.com/gettingstarted

.

Use the LabVIEW Help, available by selecting Help»LabVIEW Help in LabVIEW, to access 



information about LabVIEW programming concepts. Refer to the following locations on the 


tải về 1.5 Mb.

Chia sẻ với bạn bè của bạn:
1   ...   14   15   16   17   18   19   20   21   ...   24




Cơ sở dữ liệu được bảo vệ bởi bản quyền ©hocday.com 2024
được sử dụng cho việc quản lý

    Quê hương