Seldinger Technique
Seldinger Technique
• Method for catheterization of vessels
• Developed 1950's still popular today
• Percutaneous (through the skin) technique for arterial and venous access
• 3 vessels considered:
- Femoral-preferred site for arterial (size + accessibility)
- Brachial
- Axillary
Cannulas 
• Simple 18 guage angiographic puncture needle - one-piece open needle with a sharp beveled tip.
- Guidewire is introduced directly through the needle once the tip is fully within the bleeding vessel lumen. This 
style of needle can be used for both arterial and venous punctures.
- Guidewire is introduced directly through the needle once the tip is fully within the bleeding vessel lumen.
Guide Wires
Common guide wires


Left to right
- Straight 0.038-inch
- J-tipped 0.038-inch with introducer device( arrow) to straighten guide wire during insertion into needle hub
- Angled high-torque 0.035-inch > Angled hydrophilic-coated
0.038-inch nitinol wire with pinvise(curved arrow) for fine
control
- 0.018-inch platinum-tipped
microwire.
Basic construction of common guide wires.
1 and 2, Curved and straight safety guide wires with outer colled spring wrap, central stiffening mandril welded at 
back end only, and small safety wire (arrow) welded on inside at both ends.
3. Movable-core guide wire in which mandril can be slid back and forth and even removed completely to change 
wire stiiness, using handle incorporated into guide wire (arrow),
4. Mandril guide wire in which soft spring wrap is limited to one end of guide wire (arrow). Remainder of guide 
wire is a plain mandril. 
5, Mandril guide wire coated with hydrophilic substance (arrow).
- Guide wires are available in a number of thicknesses, lengths, tip configurations, stiffness, and materials of 
construction.
- Guide wire - the same as or slightly smaller than the diameter of the lumen at the tip of the catheter or device that 
will slide over it.


- Too big will jam, usually at the tip of the catheter.
- If a guide wire is much smaller than the end hole of the
catheter or device, there will be a gap between the guide wire and catheter that can cause vessel injury or prevent 
smooth movement over the guidewire.
- Thickness measured in one hundrethds of an inch :
0.038 Inches, 0.035 inches, 0.014 inches etc.