Pacemaker Learning Package

Asynchronous Pacing DOO 

Pacemaker will not sense the pt’s intrinsic rhythm & will pace the pt 
regardless of any intrinsic activity 

BEWARE if the mode is set in DOO or if the sensitivity number is set too 
high ( 20mV) you will cause asynchronous pacing 

Asynchronous pacing is DANGEROUS as it can cause fatal arrhythmias 
(VT & VF) 

Used in OT to prevent interference from diathermy 


Liverpool Hospital
 
Intensive Care: Learning Packages Intensive Care Unit
Pacemaker Learning Package 
LH_ICU2016_Learning_Package_Pacemaker_Learning_Package
31 | 
P a g e
 
VDD 
 

Allows for pacing in patients with ventricular pacing wires only. 

Only paces the ventricles 

Atrial and ventricular sensing 

Inhibited response to sensing in the ventricle and triggered response to 
sensing in the atrium 

Will pace the ventricle depending on what it senses in the atria and 
ventricle 

If it senses ventricular activity then it won’t pace 

If it doesn’t sense ventricular activity it will be triggered to pace the 
ventricle after it senses the hearts intrinsic atrial activity
 
Figure 19:uptodate.com.
A single pacemaker spike (marked by red lines) is followed by a QRS 
complex which is wide and bizarre and resembles a ventricular beat. 
There are some normal intervening QRS complexes that suppress the 
pacemaker when they occur faster than the rate set for the pacemaker.


Liverpool Hospital
 
Intensive Care: Learning Packages Intensive Care Unit