Pacemaker Learning Package

32 | 
P a g e
 
SENSITIVITY 

SENSITIVITY

Is the pacemaker’s ability to sense a patient’s intrinsic rhythm or 
when natural depolarisation is occurring

The sensitivity number represents the minimum size, in millivolts, 
of an electrical signal that will be detected by the pacemaker 

The higher the sensitivity (number) the less likely the pacemaker will see 
the patients rhythm 
Sensitivity
The higher the numerical value of the sensitivity, the less sensitive the 
pacemaker is.
The pacemaker looks for any intrinsic event with an amplitude greater than 
the sensitivity value.
In the example above anything over the dotted line is what the pacemaker 
see’s.
2.5
1.25
0.5
Figure 20: Temporary Cardiac Pacing.ppt P.Nekic 

Looking at the ECG trace the p waves (which represent atrial 
excitation) only have amplitude of 2-3mm. The sensitivity setting for 
atrial pacing is 0.2 -10mV. If the sensitivity setting goes above the 
amplitude of the intrinsic p wave the pacemaker will no longer be 
able to see the intrinsic p wave so then it will pace the atrium in a 
asynchronous mode, i.e it will pace the atrium regardless of the 
patients underlying rhythm


Liverpool Hospital
 
Intensive Care: Learning Packages Intensive Care Unit
Pacemaker Learning Package 
LH_ICU2016_Learning_Package_Pacemaker_Learning_Package
33 | 
P a g e
 

The same goes for the QRS complex (which represents ventricular 
contraction). The sensitivity setting for ventricular pacing is 2-20mV. 
So if the sensitivity setting goes above the amplitude of the QRS 
complex it will pace the ventricle asynchronously. 

In the above example the higher the fence the less the pacemaker sees. 
The fence, i.e. the sensitivity setting, blocks the pacemaker’s ability to see 
the patients underlying rhythm. If the fence/sensitivity is put up to the 
highest setting the pacemaker becomes asynchronous and the 
pacemaker will pace regardless 

If the sensitivity setting is set at a lower setting or the fence is lower then 
the pacemaker will see the patients underlying rhythm so the pacemaker 
will pace 

It is better to have intrinsic activity as it stimulates better contractions, and 
more effective forward flow of blood, from the chambers of the heart, 
especially when the atria and ventricles are intrinsically beating in time. 