Liverpool Hospital
Intensive Care: Learning Packages Intensive Care Unit
Pacemaker Learning Package
LH_ICU2016_Learning_Package_Pacemaker_Learning_Package
35 |
P a g e
THRESHOLDS
OUTPUT THRESHOLDS
The output threshold is the minimum amount of energy required to
“capture” the heart.
To find output threshold
If the patient is not being paced, then increase the pacemaker rate to
10 above the patient’s rate
If the patient
is already in a paced rhythm, turn the pacemaker rate
down and observe the ECG monitor for the
patients intrinsic heart
rhythm.
If the patient is still pacing at a rate of 60,
abort the checking of the
output threshold
Explain to the patient that they may briefly become ‘light-headed’
during the checking procedure.
Obtain a rhythm strip.
Decrease the output setting until loss of capture occurs.
Increase the output setting until you just see that capturing is occurring.
This is called the stimulation
Threshold.
It is usually measured in milliamps (MA) for temporary pacemakers and
volts (V) for
permanent pacemakers
You then set the output at double the threshold and add one as a
safety margin
Chia sẻ với bạn bè của bạn: