MODES OF PACING 
Over the years pacemakers have developed from the very early devices 
which merely delivered a timed impulse at set amplitude in a single 
chamber, and paid no attention to the patient’s intrinsic rhythm. 
Now even temporary pacemakers are relatively sophisticated devices which 
can deliver a range of therapeutic options from the basic fixed rate 
asynchronous, to dual chamber demand pacing. 
In order to be able to easily communicate to others the potential capabilities 
and actual programming of pacemakers, it was necessary to develop a 
code, which is shown here, generally called the NBG code. Generally the first 
three codes are utilised. 
Figure 16: NBG Code.www.cardiacrhythm.co 


Liverpool Hospital
 
Intensive Care: Learning Packages Intensive Care Unit
Pacemaker Learning Package 
LH_ICU2016_Learning_Package_Pacemaker_Learning_Package
25 | 
P a g e
 
CODES: WHAT DOES IT ALL MEAN? 

Defines the chambers that are paced/sensed 

Defines how the pacemaker will respond to intrinsic events 
 

P
ACING : first letter is chamber being 
paced
 

S
ENSING: relates to the chamber in which the pacemaker is attempting 
to 
sense
 the patient’s intrinsic rhythm.

R
ESPONSE: relates to the pacemakers 
response
to the sensed intrinsic 
beat, either triggered or inhibited 