Table of contents 1 Why is particle size important?



tải về 7.54 Mb.
Chế độ xem pdf
trang15/33
Chuyển đổi dữ liệu17.09.2022
Kích7.54 Mb.
#53206
1   ...   11   12   13   14   15   16   17   18   ...   33
Particle Guidebook 09-2019

TESTING REPRODUCIBILITY
There are currently two internationally accepted standards written on the use of 
laser diffraction: ISO 13320 (ref. 9) and USP<429> (ref. 10). Both standards state 
that samples should be measured at least three times and reproducibility must meet 
specified guidelines. Note that this means three independent measurements (i.e.
prepare the sample, measure the sample, empty the instrument, and repeat). The 
coefficient of variation (COV, or (std dev/mean)*100) for the measurement set must
be less than 3% at the D50 and less than 5% at the D10 and D90 to pass the ISO
13320 requirements. These guidelines change to less than 10% at the D50 and less
than 15% at the D10 and D90 when following the USP<429> requirements. Finally,
the guidelines all double when the D50 of the material is less than 10µm. 
While following the ISO or USP guidelines to test reproducibility is suggested, it is 
typically part of an internal specification or procedure. The specifications shown to
potential customers typically don’t include the reproducibility values. 
figure 15
|
REPORTING PSD PERCENTAGE
 
SMALLER THAN THE GIVEN SIZE
In this example, percentage of
the PSD is reported at 100nm.
figure 14
|
MEASUREMENT ERROR
Error appears exaggerated on
the Y axis because of the 
narrowness of the PSD 
12
undersize error of +/-20%
size error 
of +/-5%
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.2
0.1
20
40
60
80
100
SIZE IN 
µm
%
 UNDER
120
140



tải về 7.54 Mb.

Chia sẻ với bạn bè của bạn:
1   ...   11   12   13   14   15   16   17   18   ...   33




Cơ sở dữ liệu được bảo vệ bởi bản quyền ©hocday.com 2024
được sử dụng cho việc quản lý

    Quê hương