Troubleshooting/complications 
Failure to capture 
 
Failure to capture occurs when 

A pacer spike is present but is not followed by a corresponding 
waveform (P wave or QRS complex). 
Arrow indicates electrical stimulus without ventricular capture.
Figure 22: Critical Care Nurse.

Capture occurs when the pacemaker stimulus causes the entire chamber 
to depolarize, hopefully generating a contraction.

The ECG for failure to capture differs from the failure to fire by the 
presence of pacemaker spikes  

The pacemaker stimulus is delivered, however the patient’s heart does not 
depolarize 
 
Causes 

Inappropriate output setting 

Increased resistance to conduction 

QRS complex not visible 

Faulty cable connection 

Dislodged/fractured lead 

Tissue is refractory 


Liverpool Hospital
 
Intensive Care: Learning Packages Intensive Care Unit
Pacemaker Learning Package 
LH_ICU2016_Learning_Package_Pacemaker_Learning_Package
42 | 
P a g e
 
 
Failure to sense 
 
Failure to sense occurs 

When the generator does not detect intrinsic beats and initiates an 
inappropriate impulse that may or may not capture. 
Arrow indicates an atrial beat not sensed by the pulse generator, which is 
followed by a pacer spike. 
Figure 23: Critical Care Nurse. 
Causes 

Inappropriate sensitivity setting 

Battery depletion 

Decreased QRS voltage 

Fusion beat 

Dislodged/fractured lead 