Liverpool Hospital
Intensive Care: Learning Packages Intensive Care Unit
Pacemaker Learning Package
LH_ICU2016_Learning_Package_Pacemaker_Learning_Package
40 |
P a g e
Precautions
Do not attempt to change battery while pacing a dependent patient.
Asynchronous operation has the potential to induce life threatening
arrhythmias. Ensure a defibrillator
is at hand, and return to synchronous
operation as soon as possible
Ensure that implanted temporary pacing wires are
properly insulated
when not connected to the pacemaker, as these provide a direct
electrical connection to the myocardium….
ONLY handle the pacing wires with GLOVES
For EMERGENCIES always keep a spare
pacing box on the emergency
trolley (battery included)
Requires 9v battery
Prior to turning on check settings are appropriate
Check pacemaker mode is on
DDD on return from OT
DDD is the safest mode to use for post op cardiac patients unless
otherwise indicated